ALLERGAN (COSTA RICA) Style 20 Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled Back to Search Results
Catalog Number 20-750
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported a rupture.This record is for the right side.The device was explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported a rupture.This record is for the right side.The device was explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed broken device through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (non penetrating nicks) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Style 20 Silicone Gel Filled Breast Implant
Common Device Name
prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key25857535
Report Number9617229-2026-13651
Device Sequence Number17441487
Product Code FTR
UDI-Device Identifier10888628001794
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2014
Device Catalogue Number20-750
Device Lot Number1783286
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/26/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2026
Supplement Date Received by Manufacturer09/08/2026
Initial Report FDA Received Date07/17/2026
Supplement Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/20/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
Patient Weight121 LB
Patient RaceWhite
-
-