Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: h3, h6: no device associated with this report was received for examination.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot = > there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Patient was revised due to loosening of the tibial tray, sleeve and stem at the bone to implant interface.The patient presents with pain around the tibial implant.Lucencies are present on x-ray and the patient is scheduled for a tibial revision.Operatively, the distal femoral component is removed to gain exposure.The tibial component was well fixed at the cement interface.Once they were removed, little ingrowth was present on the sleeve.Preparation for new tibial components is performed and trialing was performed showing good rom and stability.New components are placed.No delay nor patient harm is noted.Significant bone loss was noted in the proximal tibia and could have been the reason for late loosening, as the tibial sleeve was only available with a partially porous sleeve.This would have had little contact with host bone during the previous surgery.The attune revision now has a fully porous option which was utilized during this procedure and provided more contact area.Doi: (b)(6), 2023 dor: (b)(6), 2026 affected side: right knee.
|