| Model Number |
1116-111047-21-USA-0 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device would not power up.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation.The history log from this device was reviewed for evidence of the device not powering on and any other power-related, battery, or ac faults.There are no power-related faults closer to the aware date of 06/29/2026, despite the device continuing to be used by the customer.The device returned by the customer underwent testing without duplicating the customer's report.The device was able to power on with multiple known good accessories, including the customer's returned battery (sn (b)(6)).The battery itself passed functionality testing as well and was successfully recalibrated.There was no fault found with the device.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|