| Model Number |
620-2231011-01 |
| Medical Device Problem Codes |
Failure to Deliver Shock/Stimulation (1133); Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed an unspecified "defib fault error" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed to discharge.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is reporting the evaluation of the device and correcting information submitted on the initial report.Please reference b5, h6: event description and device problem code.The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including all functional testing using known-good accessories without duplicating the report.An internal inspection of the device found no discrepancies.The multifunction cable (mfc) receptacle was replaced as a precaution.The device was recertified and returned to the customer with a recommendation to discard the mfc cable used during the event.Analysis of reports of this type has not identified an increase in trend.
|
| |
|