| Model Number |
421909 |
| Medical Device Problem Code |
Device Misassembled During Manufacturing /Shipping (2912)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation:
this complaint has been evaluated.A photograph has been received and evaluated for this complaint.Misalignment is visible in the photo.Review of the dhr showed that all relevant tests required during the manufacturing process, packaging process and final product release had been fulfilled and met the requirements.No nonconformity has been registered during the manufacturing process of the mentioned lot.No other complaint was received to this lot.The batch record review indicates no discrepancies.This issue will be monitored through the post market product monitoring review process.
|
| |
|