• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Vascutek LTD Thoraflex Hybrid Plexus

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Vascutek LTD Thoraflex Hybrid Plexus Back to Search Results
Model Number N/A
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clinical code: 4440 - thrombosis/thrombus: the site reported the event as a thrombus.Impact code: 4648 - insufficient health impact information: awaiting further information.Medical device problem code: 3190 - insufficient device problem information: awaiting further information.Component code: 4755 - part/component/sub-assembly term not applicable.Type of investigation: 4110 - trend analysis: four year review was performed for thoraflex hybrid > occlusion/thrombosis > all failure categories (excluding emboli/stenosis) jan22- jun26, this gave an occurrence rate of 0.206%.Increasing trend was identified and meeting was arranged with sme's(subject matter experts) to review this trend and discuss next steps at this time.3331 - analysis of production records: comprehensive batch review was performed, no issues were identified and the device was manufactured to required specification.Investigation findings: 3233 - results pending completion of investigation.Investigation conclusion: 11 - conclusion not yet available.
 
Event or Problem Description
Events reported from insel gruppe ag - 3064, via swissmedic report.Events reported as thrombus after frozen elephant trunk.Thrombus at the distal stent end of the frozen elephant trunk is a well-known complication.In the literature, the incidence is given as about 8-16%.Recently, we have had the impression that this complication is occurring more often in our country.The site compiled the cases of the last three months, it was found that 5 out of 12 patients had a distal thrombus, which is significantly higher than the rate described in the literature.This is significantly higher than the rate described in the literature.In addition, there was a case with serious clinical consequences last november.Despite careful intra-operative measures and established avoidance strategies on our part, thrombi formation continues to occur, which is why we would like to report this.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Thoraflex Hybrid Plexus
Common Device Name
Thoraflex Hybrid
Manufacturer (Section D)
Vascutek LTD
newmains avenue
inchinnan buiness park
renfrewshire, glasgow PA49R R
UK  PA49RR
Manufacturer (Section G)
Vascutek LTD
newmains avenue
inchinnan buiness park
renfrewshire, glasgow PA49R R
UK   PA49RR
Manufacturer Contact
laura allan
newmains avenue
inchinnan buiness park
renfrewshire, glasgow PA49R-R
UK   PA49RR
MDR Report Key25887429
Report Number9612515-2026-00042
Device Sequence Number10099440
Product Code QSK
UDI-Device Identifier05037881031750
UDI-Public05037881031750
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
P210006
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberTHP3032X100B-G
Device Lot Number25909927
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2026
Supplement Date Received by Manufacturer06/24/2026
Initial Report FDA Received Date07/21/2026
Supplement Report FDA Received Date09/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-