| Model Number |
500001002002 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 13-feb-2007 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the door failed.A field service engineer (fse) found tower door 2 was clicking during operation.Replaced the latch and verified proper door functionality.The system functioned as intended after the field service engineer repaired the device.
|
| |
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ 4000, tower door was failed.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
|
| |
|