• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATORS Back to Search Results
Model Number DDU-E2200
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key25891234
Report Number3003521780-2026-01129
Device Sequence Number9956727
Product Code MKJ
UDI-Device Identifier00815098020195
UDI-Public(01)00815098020195
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-E2200
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2026
Initial Report FDA Received Date07/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/04/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-