With all the available information, boston scientific (bsc) concludes:
device technical analysis.The watchman fxd curve? access system was discarded; therefore, returned device analysis could not be completed.Device history record review
a review was completed of the device history records (dhr) for the watchman fxd curve? access system, part # m635tu80020 batch # 0039379559.The dhr review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Labeling review:
there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Risk review:
a risk review was completed and confirmed that the events of access system - entrapped air was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to procedure.
|