| Catalog Number |
528235 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that ""[the]" staples were found jamming during use on the patient." no patient injury or harm reported.No medical intervention reported.Patient condition reported to be "fine.".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#1900141579
complaint verification testing could not be performed as no sample was returned for analysis.A verification of failure mode reported in the current manufacturing process was conducted as follows: 20 staplers were taken from the current production from p/n 528235 visistat 35w 6/box lot# 73f2600928 the staplers were functionally inspected (fired) and issue reported "misfire/jamming - info not provided" was not observed in the current manufacturing process, the staples were loaded and released correctly.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|