| Catalog Number |
SSLP-340 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported ¿exchange with no complaint against the device¿.Later healthcare professional reported ¿sitting a bit laterally¿, ¿lateral malposition¿.Capsulorraphy was performed.This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: migration ¿this report was impacted by emdr scheduled maintenance occurring (b)(6), 2026.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional/corrected data: d9, h3, h6.Device evaluation:
the device related to the reported event of migration was received on september 03, 2026, with lot number 3577799.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- migration: unable to observe this condition.None of the other observations performed during the device analysis (creases and deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿exchange with no complaint against the device¿.Later healthcare professional reported ¿sitting a bit laterally¿, ¿lateral malposition¿.Capsulorraphy was performed.This record is for the left side.Device was explanted.
|
| |
|