Model Number GIF-H170
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event is unknown.Additional information will be provided once available.The device was returned to olympus for inspection.A definitive root cause could not be identified.Based on the results of the investigation, the presumed cause is traced to component failure, as the high-frequency treatment device was used without maintaining sufficient distance from the tip.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.This mdr was submitted during the system outage from july 22, 2026 to july 27, 2026 which may result in a delay of receipt of all of the acknowledgements.
 
Event or Problem Description
The customer returned the gastrointestinal videoscope, which is an olympus asset with no reported complaint.During the evaluation of the device, c-cover burn was found.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25906241
Report Number9610595-2026-58950
Device Sequence Number8439188
Product Code FDS
UDI-Device Identifier04953170334115
UDI-Public04953170334115
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K112680
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGIF-H170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2026
Initial Report FDA Received Date07/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-