| Model Number |
GIF-H170 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided once available.The device was returned to olympus for inspection.A definitive root cause could not be identified.Based on the results of the investigation, the presumed cause is traced to component failure, as the high-frequency treatment device was used without maintaining sufficient distance from the tip.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.This mdr was submitted during the system outage from july 22, 2026 to july 27, 2026 which may result in a delay of receipt of all of the acknowledgements.
|
| |
|
Event or Problem Description
|
|
The customer returned the gastrointestinal videoscope, which is an olympus asset with no reported complaint.During the evaluation of the device, c-cover burn was found.There was no patient involvement.
|
| |
|