| Model Number |
CU-172R |
| Medical Device Problem Codes |
Display Difficult to Read (1181); Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Details of complaint:
the biomedical engineer (bme) reported that the g7 monitor had vertical lines going across the entire display.The bme also stated that the display would eventually dim off and the screen remains black.No patient harm was reported.Investigation summary:
the device was returned to nihon kohden for evaluation and repair.However, the reported issue could not be replicated, and the device was returned to the customer without repair.Since the complaint could not be replicated, a definitive cause for the event cannot be established.A possible explanation for the event is a transient software issue.A review of the device's complaint history by serial number reveals no similar issue.Nihon kohden will continue to trend and monitor.Additional information:
b4 date of this report
d9 device available for evaluation
g3 date received by manufacturer
g6 type of report
h2 if follow-up, what type?
h3 device evaluated by manufacturer
h6 event problem and evaluation codes
h11 additional manufacturer narrative.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported that the g7 monitor had vertical lines going across the entire display.The bme also stated that the display would eventually dim off and the screen remains black.No patient harm was reported.
|
| |
|