• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL Back to Search Results
Catalog Number BXB093901A
Medical Device Problem Code Device Endoleak (4074)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 02/10/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of manufacturing records verified that the lot involved in this complaint met all pre-release specifications.The device remains implanted and was, therefore, not available for analysis.No clinical images enabling direct assessment of product performance were returned for evaluation.Cause of the reported event cannot be established based on the information reported to gore.Further information regarding this event was requested by gore, but no further information has been reported, therefore this investigation is considered complete.Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following information was reported through the aaa 24-01 tambe post approval study: on (b)(6) 2026, the patient underwent endovascular procedure to treat a thoracoabdominal aortic aneurysm using a gore® excluder® thoracoabdominal branch endoprosthesis (tambe) device system and gore® viabahn® vbx balloon expandable endoprosthesis (vbx device).On (b)(6) 2026, a type 1c endoleak was identified in the celiac artery.On (b)(6) 2026, a reintervention was performed to treat the type 1c endoleak by extending with another vbx device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Common Device Name
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
nick
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key25906527
Report Number2017233-2026-07813
Device Sequence Number12210192
Product Code PRL
UDI-Device Identifier00733132658794
UDI-Public(01)00733132658794(17)280202
Combination Product (Y/N)Y
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P160021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberBXB093901A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2026
Initial Report FDA Received Date07/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-