Model Number MODEL 250
Device Problem Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2013
Event Type  malfunction  
Event Description
It was reported that the physician's handheld was at the align screen and keeps going "around and around".Troubleshooting was performed by a manufacturer employee.The battery was removed and reinserted, a hard reset was performed; however, the screen still continues to go "around and around".The edges of the screen were cleaned with a card and this did not resolve the issue.A new programming tablet was sent to the physician.The handheld is expected to be returned, but has not been received to date.
 
Manufacturer Narrative
Device manufacturing records were reviewed.Review of manufacturing records of the handheld confirmed all quality tests were passed prior to distribution.
 
Event Description
The handheld device was returned to the manufacturer for analysis.Analysis of the returned handheld found that the cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Once the display was replaced with a known good display, no further anomalies associated with the handheld performance were identified during the analysis.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Manufacturer Narrative
Device failure occurred, but did not cause or contribute to death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3577154
MDR Text Key4074439
Report Number1644487-2014-00245
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/19/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number840143
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received04/15/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-