Model Number 300-20
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/25/2014
Event Type  malfunction  
Event Description
It was reported that the vns patient¿s device was tested and diagnostic results revealed high lead impedance (dc dc ¿ 7).The patient was last seen in (b)(6) 2013 and diagnostic results showed lead impedance within normal limits (dc dc ¿ 1) at the time.There were no reports of falls, adverse events, or contributory events that could have led to patient's high impedance.The neurologist elected not to have the patient¿s device programmed off.X-rays were taken and were reported to show no lead discontinuity.Replacement surgery was scheduled but canceled per the patient¿s request.
 
Manufacturer Narrative
Device manufacturing records were reviewed.Review of manufacturing records confirmed that the lead passed all functional tests prior to distribution.Device failure is suspected, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOL LEAD MODEL 300
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3758923
MDR Text Key21727970
Report Number1644487-2014-01055
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/31/2003
Device Model Number300-20
Device Lot Number1934
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received03/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/17/2001
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age25 YR
-
-