• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS INC BIPOL LEAD MODEL 300

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS INC BIPOL LEAD MODEL 300 Back to Search Results
Model Number 300-20
Device Problems High impedance (1291); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/10/2014
Event Type  malfunction  
Event Description
It was reported that during generator replacement surgery, high impedance was observed and the patient underwent both generator and lead replacement.Only the explanted generator was returned for analysis.Analysis of generator was completed.The end of service condition was the result of normal, expected battery depletion based on the battery life calculation, the electrical test results and the bench evaluation.The device performed according to functional specifications.Analysis of the generator in the pa lab concluded that no abnormal performance or any other type of adverse condition was found.The physician's office reported that the physician would not remember any further information; therefore, no additional relevant information was able to be obtained.
 
Manufacturer Narrative
Device failure is suspected, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOL LEAD MODEL 300
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3778328
MDR Text Key4356845
Report Number1644487-2014-01132
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2003
Device Model Number300-20
Device Lot Number2802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received04/01/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/17/2001
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age42 YR
-
-