Catalog Number AR-1662BC-8
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Code Available (3191)
Event Date 03/20/2014
Event Type  Injury  
Event Description
It was reported that the forked tip portion broke off in the patient upon bone insertion.This portion of the tip was not retrieved from patient.Ar-1662bc-8 was no longer sterile and the item was discarded.Procedure was a bicep tenodesis.
 
Manufacturer Narrative
No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.If the device is returned and additional information is obtained, a follow-up report will be submitted.The complainant's event is typically caused by not inserting the implant co-axial to the bone tunnel, prying/leveraging or impacting the eyelet against hard bone.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.Device discarded by the facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIVELOCK TENODESIS BIO- COMPOSITE 8 X 19.5MM
Type of Device
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer Contact
vik bajnath, sr. mdr analyst.
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key3782683
MDR Text Key4378838
Report Number1220246-2014-00067
Device Sequence Number1
Product Code MAI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101823
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 04/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/01/2015
Device Catalogue NumberAR-1662BC-8
Device Lot Number837107
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/07/2014
Initial Date FDA Received05/01/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
-
-