It was reported that the crux vcf deployment was successfully completed.Upon removing the delivery system, the physician tightened the hemostasis valve leaving the tracking tip extended per ifu.Upon pulling the delivery catheter out of the body, the edge of the tracking tip caught on the filter web causing the filter to move caudally with the delivery catheter removal.The result of this movement was that the filter was now positioned low in the ivc.It was reported that the caudal retrieval tail was now positioned at the confluence of the right iliac vein.The physician chose to retrieve the crux vcf because of the low positioning.The retrieval was completed successfully.The physician then deployed a second crux vcf.The filter was deployed in good position.The delivery catheter was removed from the body without incident.There was no surgical intervention required.
|