• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Device Stops Intermittently (1599); Difficult or Delayed Activation (2577)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2014
Event Type  malfunction  
Event Description
It was reported that the nurse¿s handheld device was not working properly.The handheld device was slow and worked intermittently.The programming wand was recently replaced and suspected to be contributing to the reported issues.The handheld device took a long to interrogate some devices but worked well with others.The handheld device sometimes failed to interrogate.Troubleshooting was performed and determined that the handheld device needed to be replaced.The handheld device has not been returned to date.
 
Event Description
The handheld and flashcard were received for analysis.Analysis of the handheld was completed on (b)(4) 2014.An analysis on the returned handheld found that the device would not power on using a known good ac adapter.The cause for the anomaly is associated with damaged solder connections to the pcb.Because of the damage, the handheld received power from the ac adapter intermittently.The cause for the damage to the solder connections is most likely associated with mishandling of the device as it appears the connector detents are not being compressed when removing or manipulating the serial data cable, thus placing an extensive amount of force on the pcb and attached cable receptacle.Analysis of the flashcard was completed on (b)(4) 2014.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Manufacturer Narrative
Device manufacturing records were reviewed.Review of manufacturing records confirmed that the handheld device passed all functional tests prior to distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3847263
MDR Text Key4408579
Report Number1644487-2014-01408
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063525
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received06/09/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-