Analysis of the generator and lead was completed.In the analysis lab, the device output signal was monitored for more than 24-hrs, while the generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator's output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.The pulse generator diagnostics were as expected for the programmed parameters.In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no performance or any other type of adverse conditions found with the pulse generator.Note that a large portion of the lead assembly (body) including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis, three broken coil strands were observed on the (-) connector pin quadfilar coil approximately 120mm-122mm from the end of the connector boot.Scanning electron microscopy was performed identified the area as having evidence of being worn to the point of fracture with flat spots and pitting on the coil surface.During the visual analysis the (+) connector ring quadfilar coil appeared to be broken approximately 119mm and 131mm from the end of the connector boot.Scanning electron microscopy was performed on the (+) connector ring quadfilar coil break (found at 119mm) and identified the area as being mechanically damaged which prevented identification of the coil fracture type with pitting.Flat spots and pitting were observed on the coil surface.Scanning electron microscopy was performed on the (+) connector ring quadfilar coil break (found at 131mm) and identified the area as having evidence of being worn to the point of fracture with flat spots on the coil surface and no pitting.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.Low magnification sem analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution.The abraded openings found on the outer and inner silicone tubes most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids found inside the outer and inner silicone tubes.With the exception of the observed discontinuities, the condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.Continuity checks of the returned lead portions were performed, during the visual analysis, and no other discontinuities were identified.Note that a break of a few strands would still allow current flow through that portion of the lead.This observation is supported by results of electrical measurements which verified continuity between the ends of this lead section.Note that since a large portion of the lead assembly (body) including the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.
|