CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/11/2014
Event Type  malfunction  
Event Description
The handheld and flashcard were received for analysis.Analysis of the handheld was completed on (b)(4) 2014.During the analysis it was identified that the handheld was received with the lock button in the locked position.Since the lock button was locked, the handheld buttons and touchscreen were unresponsive.Once the lock button was moved to the unlocked position no anomalies associated with the handheld performance were noted during testing using the ac adapter or the main battery with a full charge.Analysis of the flashcard was completed on (b)(4) 2014.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Event Description
Additional information was received stating that the hospital staff was well trained in proper usage of the handheld.The handheld device was programmed off and intentionally returned with the lock button engaged.
 
Event Description
Additional information was received clarifying that the handheld device had a frozen screen and was not responding to the user.The handheld device is expected to be returned to the manufacturer for analysis but has not been returned to date.
 
Manufacturer Narrative
Lot #, corrected data: previously submitted mdr incorrectly reported this information.
 
Event Description
It was reported that the physician's handheld does not respond to any command and has self behave.It is unknown if troubleshooting was performed.Attempts to obtain additional relevant information have been unsuccessful to date.
 
Manufacturer Narrative
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3892490
MDR Text Key4723289
Report Number1644487-2014-01581
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 05/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063833
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received09/07/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-