• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; APS 1 (HEART LUNG CONSOLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; APS 1 (HEART LUNG CONSOLE) Back to Search Results
Model Number 801763
Device Problems Bent (1059); Device Stops Intermittently (1599)
Patient Problem Not Applicable (3189)
Event Date 04/08/2014
Event Type  malfunction  
Event Description
The field service representative (fsr) reported that during preventative maintenance (pm) of the device, the power cable of the perfusion system had a pin that was bent.This unit would work intermittently with the bent pin.Since the event occurred during preventative maintenance, there was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed by the field service representative (fsr).The fsr could not bend the pin back in the power cable.The fsr replaced the power cable and completed the preventative maintenance (pm).System-1 and pumps met manufacturer's specifications before being returned to clinical use.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
APS 1 (HEART LUNG CONSOLE)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key3900176
MDR Text Key4648733
Report Number1828100-2014-00322
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 04/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/16/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/08/2014
Initial Date FDA Received05/01/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-