• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - FREMONT (SUD) OPTICROSS?; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - FREMONT (SUD) OPTICROSS?; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H749518080
Device Problem Difficult To Position (1467)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/17/2014
Event Type  malfunction  
Event Description
Reportable based on analysis completed on (b)(4) 2014.It was reported that the transducer did not move back to its initial position.The opticross¿ imaging catheter was used to visualize an unspecified lesion.It was noted during percutaneous coronary intervention, post intravascular ultrasound, the transducer did not move back to its initial position.The procedure was completed with another of the same device.No patient complications were reported and the patient's condition is good.However, returned device analysis revealed an open hole at the sheath lap joint assembly.
 
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the complaint device was returned for evaluation.The telescope assembly was not able to properly pull back, advance, or retract.An open hole was observed at the sheath lap joint section of the device.Fluid was leaking from the open hole at the sheath lap joint assembly when the catheter was flushed.During image characterization testing, no image appeared in the system due to electrical open at proximal.No imaging core windup was found within the telescope section of the device.No other issues or defects were observed during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS?
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (SUD)
47215 lakeview blvd phone
west dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (SUD)
47215 lakeview blvd phone
west dock
fremont CA 94538
Manufacturer Contact
ingrid matte
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key3909743
MDR Text Key15119375
Report Number2134265-2014-03778
Device Sequence Number1
Product Code OBJ
Combination Product (y/n)N
PMA/PMN Number
K123621
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 06/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/24/2015
Device Model NumberH749518080
Device Catalogue Number51808
Device Lot Number16767390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/06/2014
Initial Date FDA Received07/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/25/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-