• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. BIPOL LEAD MODEL 300

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. BIPOL LEAD MODEL 300 Back to Search Results
Model Number 300-20
Device Problems Device Inoperable (1663); Malposition of Device (2616)
Patient Problem Neck Pain (2433)
Event Date 07/24/2014
Event Type  Injury  
Manufacturer Narrative
Corrected data: the initial mfr.Report inadvertently reported the event as both serious injury and malfunction; however, this should only have been reported as a serious injury.
 
Event Description
It was reported that the vns patient was experiencing intense pain in her neck.The patient¿s device was subsequently disabled; however, the patient continued to experience stabbing pains but less frequently than before disablement.X-rays were taken and the lead reportedly appeared to be ¿unwrapped from the nerve.¿ the patient underwent generator and lead replacement surgery on (b)(6) 2014.The explanting facility discarded the explanted lead; therefore, no analysis can be performed.
 
Event Description
Additional information was received which stated the patient experienced neck pain prior to the revision surgery, even though the diagnostics were within limits.It was noted that when the lead was explanted fluid was found inside.After the revision surgery, the patient's neck pain didn't go away for the first few days after the surgery, but then it did resolve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOL LEAD MODEL 300
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4027949
MDR Text Key12201812
Report Number1644487-2014-02075
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup
Report Date 07/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2002
Device Model Number300-20
Device Lot Number255
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/27/2014
Initial Date FDA Received08/21/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received09/29/2014
11/20/2014
10/31/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2000
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age19 YR
-
-