Model Number MODEL 250
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Information (3190)
Event Date 07/31/2014
Event Type  malfunction  
Event Description
Analysis of the returned handheld device and software flashcard was completed.During analysis, it was identified that the handheld was unable to advance past the welcome screen.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Once the display was replaced with a known good display, no further anomalies associated with the handheld performance were identified during the analysis.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Manufacturer Narrative
Device failure occurred, but did not cause or contribute to death or serious injury.
 
Event Description
It was reported that the neurologist was unable to pass the alignment screen on her handheld device.Multiple hard resets were performed, the screen was cleaned, and the lock button was confirmed to be not engaged.However, the issue did not resolve.The handheld device was returned to the manufacturer where analysis is currently underway.
 
Manufacturer Narrative
Device manufacturing records were reviewed.Review of manufacturing records confirmed that the handheld device passed all functional tests prior to distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4037865
MDR Text Key4785960
Report Number1644487-2014-02105
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/31/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063830
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 09/10/2014
Initial Date FDA Received08/26/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/07/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-