Model Number MODEL 250
Device Problem Output Problem (3005)
Patient Problem Neck Pain (2433)
Event Date 02/20/2014
Event Type  malfunction  
Event Description
It was reported that the patient's generator was found to be programmed to 1.5ma output current at 180 minutes off postoperatively which caused pain in the neck for the patient.It was reported that the generator was not deliberately programmed to these settings.It was reported that the device was programmed off and the pain subsided.The generator is now programmed on and the patient is doing well.It was reported that two patients were implanted this same day and it is believed that a cross-programming event occurred.
 
Event Description
Additional information was received stating that the vns patient did not experience any lasting symptoms from this event.Additional programming and diagnostic history was provided; however, it did not show the cross programming event and could not be fully assessed due to incorrect timestamps of the programming events.
 
Manufacturer Narrative
Analysis of programming history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4118480
MDR Text Key4837012
Report Number1644487-2014-02469
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063525
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/16/2014
Initial Date FDA Received09/25/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/14/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age20 YR
-
-