It was reported that the vns patient underwent generator replacement surgery on (b)(6) 2014 due to end of service.Prior to replacement, the surgeon stated that some sort of silicone based substance was leaking from the generator.The substance was described as white in color, and the surgeon believed that it was an inorganic-based substance.A sample of this was sent to the pathology lab at the hospital for analysis.No results of the pathology lab¿s analysis have been received to date.The generator was explanted and returned to the manufacturer where analysis is currently underway.
|
Analysis of the generator was completed on 10/31/2014.Residue was observed on the pulse generator feed-thru assembly and is known to be related to the manufacturing process of the component.There was evidence of body fluid remnants on the case exterior.The septum was cored, but no evidence of dried fluids was found in the septum cavity.No leaking white substance was observed on the pulse generator.The module performed according to functional specifications.There was no condition noted during the product analysis evaluation that would suggest any anomaly with the device.The final pathology results indicate soft tissues, left side of chest, organized fat necrosis.
|