Model Number MODEL 250
Device Problems Failure to Charge (1085); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2014
Event Type  malfunction  
Event Description
It was reported that the physician's handheld was not holding a charge and would shut down when unplugged from the electrical outlet.A new programming tablet was provided to the physician.The handheld was received for analysis.Analysis is underway, but has not been completed to date.
 
Event Description
Analysis of the handheld was completed on 11/17/2014.During the analysis the main battery was able to hold a change and power the handheld for over an hour.During the analysis it was identified that the handheld would not power on.The cause for the anomaly is associated with a swollen main battery.The swollen battery bent the battery cover causing the battery latch switch to register that the latch was unlocked, thus preventing the handheld from powering on (this condition can also cause the handheld to power off intermittently).No further anomalies were identified.Analysis of the flashcard was completed on 11/17/2014.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Manufacturer Narrative
Device failure occurred, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4170973
MDR Text Key17994411
Report Number1644487-2014-02683
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1073167
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/23/2014
Initial Date FDA Received10/14/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/17/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-