Model Number MODEL 250 |
Device Problems
Failure to Charge (1085); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/18/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that the physician's handheld was not holding a charge and would shut down when unplugged from the electrical outlet.A new programming tablet was provided to the physician.The handheld was received for analysis.Analysis is underway, but has not been completed to date.
|
|
Event Description
|
Analysis of the handheld was completed on 11/17/2014.During the analysis the main battery was able to hold a change and power the handheld for over an hour.During the analysis it was identified that the handheld would not power on.The cause for the anomaly is associated with a swollen main battery.The swollen battery bent the battery cover causing the battery latch switch to register that the latch was unlocked, thus preventing the handheld from powering on (this condition can also cause the handheld to power off intermittently).No further anomalies were identified.Analysis of the flashcard was completed on 11/17/2014.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
|
|
Manufacturer Narrative
|
Device failure occurred, but did not cause or contribute to a death or serious injury.
|
|