Additional information was received that product analysis was completed on the generator and lead.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery, 2.941 volts as measured shows an ifi=no condition.The data in the diagaccumconsumed memory locations revealed that 57.124% of the battery had been consumed.There were no performance or any other type of adverse conditions found with the pulse generator.Note that a portion of the (-) green electrode quadfilar coil was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis of the returned 455mm portion the end of the (-) connector pin quadfilar coil appeared to be broken at the distal end of the anchor tether.Scanning electron microscopy was performed and identified the area as being mechanically damaged which prevented identification of the coil fracture type with pitting.Pitting was observed on the coil surface.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.Low magnification sem analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution.The abraded opening and slice mark found on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.With the exception of the observed discontinuity the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no other discontinuities were identified.Based on the findings in the product analysis lab, there is evidence to suggest a discontinuity in the returned portions of the device which may have contributed to the stated allegation of 'fracture of 'high impedance' note that since a portion of the (-) green electrode quadfilar coil was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.
|