During review of the in-house programming/diagnostic history database, it was observed that during interrogation on office visit on (b)(6) 2010 the patient's settings were different than what were programmed at the same office visit.The settings found were indicative of a faulted diagnostic test; however, review of system diagnostic testing from office visit on (b)(6) 2010 was within normal limits.The physician corrected the settings; however, the magnet pulse width was not corrected.The device was interrogated prior to the patient leaving the office on (b)(6) 2010 as recommended by device manufacturer to ensure the device is at the correct settings; however, the physician did not correct the magnet pulse width back to the previous setting.No patient adverse events were reported.
|