The patient underwent a full replacement on (b)(6) 2016 due to battery depletion, unable to interrogate due to battery depletion and lead discontinuity.The generator and lead were received for analysis on 06/13/2016.Product analysis for the generator was completed and approved on 06/22/2016.An open can measurement of the battery voltage confirmed that the eri flag had been properly set; the device was operating at a low battery condition (depleted).Based on the bench analysis and the electrical test results, the device exhibited current consumption rates that were within specification; thereby, demonstrating normal battery depletion to an end of service condition.With the exception of the parameters that are associated with a low battery condition, the device performed according to functional specifications.Therefore, the electrical performance of the generator, as measured in the pa lab, will be used to conclude that no abnormal performance or any other type of adverse condition was found with the generator.
|