Product analysis was completed on the lead on (b)(6) 2014.The (+) white and (-) green electrodes were not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis of the returned 125mm portion the end of the (-) green electrode tri-filar coil appeared to be broken approximately 15mm from the distal end of the anchor tether.Scanning electron microscopy was performed and identified the area as being mechanically damaged which prevented identification of the coil fracture type.Fine pitting was observed on two of the broken coil strands.Evidence of a stress induced fracture (torsional appearance) which most likely completed the fracture was observed on the remaining broken coil strand.Pitting and flat spots were observed on the coil surface.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.Low magnification sem analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution.With the exception of the observed discontinuity the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no other discontinuities were identified.Product analysis was completed on the generator on 12/22/2014.An end-of-service warning message was verified in the pa lab.With the pulse generator case removed and the battery still attached to the pcb, the battery measured 1.971 volts, indicating an eos condition.The data from the generator shows an increase of impedance to 9434 ohms, which occurred on (b)(6) 2014.There were no performance or any other type of adverse conditions found with the pulse generator.
|