• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C0430
Device Problems Detachment Of Device Component (1104); Difficult To Position (1467); Physical Resistance (2578)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/10/2014
Event Type  malfunction  
Event Description
The patient was undergoing a coil embolization procedure using ruby coils.During the procedure, the physician met resistance while advancing a ruby coil.Upon retraction, the ruby coil unintentionally detached within the microcatheter.The microcatheter was removed with the detached ruby coil inside.The physician had difficulty positioning a new ruby coil and it unintentionally detached.A snare device was required to successfully remove the detached ruby coil from the body.There was no report of an adverse effect on the patient.
 
Manufacturer Narrative
Conclusion: the product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This mdr is associated with mdr 3005168196-2014-00854.The hospital discarded the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
1351 harbor bay parkway
alameda CA 94502
Manufacturer (Section G)
PENUMBRA, INC.
1351 harbor bay parkway
alameda CA 94502
Manufacturer Contact
kathleen kidd
1351 harbor bay parkway
alameda, CA 94502
5107483200
MDR Report Key4317875
MDR Text Key22155213
Report Number3005168196-2014-00853
Device Sequence Number1
Product Code HCG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120330
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date06/05/2018
Device Catalogue NumberRBY2C0430
Device Lot NumberF35855
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/10/2014
Initial Date FDA Received12/10/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/05/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-