During generator replacement surgery, it was reported that this replacement generator was not used because it ¿did not work.¿ additional information was received that an error message of ¿checksum error¿ was received on the vns programming system when communicating with the generator.This error message is a communication issue message.The surgeon elected to discard the generator and discard it due to the check sum error message.After the wand 9v battery was replaced, the programming system worked as intended.However, the wand battery was not replaced until after the generator was discarded.Following the replacement of the wand battery, a new generator was placed and interrogated successfully.The generator was received by the manufacturer for analysis.However, analysis has not been completed to date.
|