The insulin pump passed all functional testing including the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.The device had a cracked case at the display window corner, cracked reservoir tube lip, minor scratched lcd window, cracked lcd window and cracked belt clip slot.Data analysis customer daily insulin total average from 61.2 to 108.95u, total basal insulin from 36.75 to 61.2 and total bolus insulin 0 to 67.0u.The insulin pump passed the delivery volume accuracy test.
|