Device evaluation: device evaluation of electrode belt sn (b)(4) was completed.The reported problem (gel not deployed) was confirmed.Upon investigation the rear therapy electrodes properly deployed gel but the front therapy electrode did not deploy gel.The cause for the gel deployment failure was a leak in the seal surrounding the therapy electrode gas generator o-ring.The leak allowed the gas from the gas generator to escape through the leak instead of forcing gel out of the exit ports of the therapy electrode.The root cause for the o-ring leak could not be positively identified.The lifevest electrode belt therapy electrodes have a surface area well above (b)(4) requirements.The larger surface area helps to limit the risk of a patient receiving a burn if the electroconductive gel does not fully deploy.Additional inappropriate defibrillation narrative: inappropriate defibrillations are an anticipated risk associated with the use of the lifevest.Patients are instructed through alarms, voice messages, ifu, and training to press the response buttons to prevent an inappropriate defibrillation.(b)(4).
|