Model Number 5392 |
Device Problem
Positioning Problem (3009)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/04/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that the seal on the external pulse generator (epg) was not in place.It is expected that the epg will be returned to the manufacturer for servicing.There was no patient involvement.
|
|
Manufacturer Narrative
|
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
|
|
Manufacturer Narrative
|
Product event summary: analysis confirmed the customer comment that the seal was not in place, the main seal was pinched and exposed outside of the case.The main seal was inspected and no anomalies were observed but it was to be sent to failure analysis as a preventive.It was additionally noted that the output connector assembly black wire was exposed at the crimp and that during removal for repair the black came loose from the crimp, that the display wire insulation was pinched though the wire insulation was not compromised and that the encoder nuts were not torqued to specification.All found defective parts were replaced and all other identified issues were resolved.
|
|
Event Description
|
It was further reported that the generator was returned for service, and analyzed.
|
|
Manufacturer Narrative
|
Further analysis was performed on the patient cable assembly.Visual inspection revealed that the black wire was pulled from the crimp.Conclusion: the wire was pulled from the crimp at service, two of the seven wires appear to not have been included in the crimp which would cause less pressure in the crimp.
|
|
Search Alerts/Recalls
|