Model Number 5392
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Patient Involvement (2645)
Event Date 12/04/2014
Event Type  malfunction  
Event Description
It was reported that the external pulse generator (epg) displayed an error and the device was "not recoverable." the epg has been returned for servicing.There was no patient involvement.
 
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
Product event summary: analysis confirmed the reported event, the main printed circuit board (pcb) was out of electrical specification.It was also noted that the display wires were pinched.
 
Manufacturer Narrative
Further analysis was performed on the main board.Visual inspection revealed no anomalies.Bench analysis found that all tests passed.The log was also analyzed.The device had a normal power-up, during a scheduled battery/super cap measurement the super cap measured high which caused an error.The next power up cycle caused the displayed error to occur.Conclusion: it was confirmed by log analysis that an error for super cap high triggered out of service error, however the error was not reproducible in analysis.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4497506
MDR Text Key5334215
Report Number3004593495-2015-00006
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 12/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-