Model Number 304-20 |
Device Problem
High impedance (1291)
|
Patient Problem
No Information (3190)
|
Event Date 01/03/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that device diagnostics resulted in high impedance (6592 ohms).Previous device diagnostics on (b)(6) 2015 were within normal limits (2736 ohms).X-rays were taken and did not identify any obvious discontinuities with the vns system.The generator was programmed off and the patient was referred to surgeon.No known surgical interventions have occurred to date.
|
|
Manufacturer Narrative
|
Device failure is suspected, but did not cause or contribute to a death or serious injury.
|
|
Event Description
|
Additional information was received that the patient is undergoing ent procedures to treat painful stimulation in the neck and head, voice alteration, and coughing that he has been experiencing since early january.However, the device being disabled has also helped diminish these symptoms as they were thought to be related to stimulation during high impedance.No known vns surgical interventions have occurred to date.
|
|
Manufacturer Narrative
|
Device failure is suspected, but did not cause or contribute to a death.
|
|
Event Description
|
Additional information was received that the patient underwent a recent brain surgery and the vns is still disabled.The patient's previous neurologist has no additional information to provide.No known surgical interventions have occurred to date.
|
|