• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Patient involvement (2648)
Event Date 01/02/2015
Event Type  malfunction  
Event Description
Sorin group (b)(4) received a report that the s5 gas blender displayed an error message and the gas flow rate could not be adjusted above 1.8 lpm during set up.There was no patient involvment.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the s5 gas blender displayed an error message and the gas flow rate could not be adjusted above 1.8 lpm during set up.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Manufacturer Narrative
As a result of the investigation, sorin group (b)(4) stated that there was a defective co2 and o2 flow control and measuring bridge.The defective parts were replaced and the gas blender was tested without any issues.No further investigation is required and no trend has been identified.No nonconformities were noted during manufacturing record review.The issue will be monitored for trends and if identified, corrections will be recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munchen D 809 39
GM  D 80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munchen D 809 39
GM   D 80939
Manufacturer Contact
cheri voorhees,mgr
14401 w. 65th way
arvada, CO 80004
3034676527
MDR Report Key4546784
MDR Text Key5522031
Report Number9611109-2015-00029
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/26/2015
Initial Date FDA Received02/20/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/11/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-