Model Number MODEL 250
Device Problems Loose or Intermittent Connection (1371); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/09/2015
Event Type  malfunction  
Event Description
It was reported that the physician's handheld charging cable is loose and a fault message was received.The physician was provided a new programming tablet.The handheld and cables were received for analysis.Analysis is underway, but has not been completed to date.
 
Event Description
Product analysis was completed for the handheld device.During the analysis, it was identified that the handheld would not power on using the ac adapter.The cause for the anomaly is associated with damaged sync connector leads and broken sync connector solder connections to the pcb.Because of the damage, the handheld was unable to receive power from the ac adapter.The cause for the damage to the connector leads and solder connections is most likely associated with mishandling of the device as it appears the connector detents are not being compressed when removing or manipulating the serial data cable, thus placing an extensive amount of force on the pcb and attached cable receptacle.No other anomalies were identified.Product analysis was completed for the software flashcard.No anomalies associated with flashcard software were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4586145
MDR Text Key5336410
Report Number1644487-2015-04103
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1075852
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received03/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-