An analysis was performed on the returned lead portions.A portion of the lead was not returned to the manufacturer and therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis of the lead, the quadfilar coil appeared to be broken at the proximal end of the anchor tether.Analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution.An abraded opening was found on the outer silicone tubing.No other obvious anomalies were noted.
|