Model Number 300-20 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/27/2015 |
Event Type
malfunction
|
Event Description
|
During a generator replacement on (b)(6) 2015 due to battery depletion, the surgeon identified the lead insulation on the electrode to be somewhat discontinuous within the generator pocket.Multiple areas of the electrode lead insulation were discontinuous with exposed wire.The surgeon therefore decided to replace the lead along with the generator.The explanted generator and lead were received by manufacturer on 3/12/2015.Analysis is underway but has not been completed.
|
|
Event Description
|
The pulse generator diagnostics were as expected for the programmed parameters.The battery shows an ifi=no condition.There were no performance or any other type of adverse conditions found with the pulse generator.The lead assembly was returned in four portions with one tie down attached.The electrodes were not returned.Multiple abraded opening were observed on the outer silicone tubing and on the marked and unmarked inner silicone tubing.The outer silicone tubing appeared to be compressed in some areas.
|
|
Manufacturer Narrative
|
Device failure is suspected, but did not cause or contribute to a death or serious injury.
|
|
Manufacturer Narrative
|
Device failure occurred, but did not cause or contribute to a death or serious injury.
|
|
Search Alerts/Recalls
|