CYBERONICS INC PULSE GEN MODEL 103 GENERATOR Back to Search Results
Model Number 103
Event Date 02/23/2015
Event Type  Malfunction  
Event Description

It was reported that the vns patient¿s device showed an ifi condition after being implanted for approximately two years and the patient was experiencing an increase in seizures. The ifi condition is believed to be premature battery depletion. A battery life calculation using the available programming history showed approximately 1. 3 years remaining. Attempts for additional relevant information have been unsuccessful to date.

 
Manufacturer Narrative

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePULSE GEN MODEL 103
Type of DeviceGENERATOR
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key4638131
Report Number1644487-2015-04271
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation
Type of Report Initial
Report Date 03/04/2015
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received03/27/2015
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date10/31/2014
Device MODEL Number103
Device LOT Number202321
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Event Location Other
Date Manufacturer Received03/04/2015
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured11/19/2012
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-