• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T167
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Atrial Fibrillation (1729)
Event Date 02/03/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.Product analysis could not confirm the reported allegation as all functions of the device are within specifications.The pacing, sensing, and shocking functions were verified per bench top testing.The device functioned normally.
 
Manufacturer Narrative
(b)(4).As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that the patient with this implantable cardioverter defibrillator (icd) received inappropriate shocks for atrial fibrillation (af) which resulted in therapy exhaustion.Boston scientific technical services (ts) discussed programming changes.The device was then reprogrammed to monitor only (mo) zone off, ventricular tachycardia (vt) 200 bpm with the rhythm id on.This icd remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key4669148
MDR Text Key13976220
Report Number2124215-2015-02815
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/16/2007
Device Model NumberT167
Other Device ID NumberVITALITY 2 DR EL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2015
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/20/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0185; T167; 4086
Patient Age38 YR
-
-