Model Number 229047
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Code Available (3191)
Event Date 01/11/2015
Event Type  Injury  
Event Description
It was reported the programmer was not working.It was also reported that the analyzer stopped working during a case (not during lead testing).It was noted that an error screen appeared causing the programmer to need rebooting.A second / spare programmer was used for the rest of the procedure.The programmer and analyzer were returned for repair.No patient complications were reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis found the analyzer back up battery had expiration date that had expired and the battery voltage was low.The analyzer could not stay on by itself due to the low battery voltage.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4676462
MDR Text Key5673708
Report Number2182208-2015-00838
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number229047
Device Catalogue Number229047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/11/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
2090 PROGRAMMER