Analyses of the returned generator and lead have been completed.Analysis concluded proper functionality of the generator and that no abnormal performance or any other type of adverse condition was found.The positive coil was found to be protruding out (still within the outer tubing) from the inner silicone tubing at what appear to be abraded/tear openings.Also, an abraded opening was noted on the inner tubing of the negative coil.Note that since a portion of the lead (including the electrode array) was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portion.
|