Product analysis (pa) for the returned generator was completed.A comprehensive automated electrical evaluation showed the generator performed according to functional specifications, with the exception of the test "backup cap neg to can", which has already been captured in mfr.Report 1644487-2017-03534.There were no additional performance or any other type of adverse conditions found with the generator.Pa for the returned lead was completed.Allegations of lead fractures were confirmed.The electrodes were not returned for analysis; therefore, a complete evaluations could not be performed on the entire lead product.During the visual analysis, the outer silicone tubing appeared to be twisted and compressed in several areas.Fractures were identified and scanning electron microscopy found that they were stress induced fractures.Additionally, pitting was observed on the surface of some of the breaks.Abraded openings found on the outer silicone tubing most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.What appeared to be white deposits were observed in various locations.Energy dispersion spectroscopy was performed on one of the deposits and identified the deposit as containing silicon, phosphorus, sodium, magnesium, and calcium.With the exception of the observed discontinuities and the abraded openings, the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.Based on the findings in the pa lab, there is evidence to suggest discontinuities in the returned portion of the device, which may have contributed to the stated allegations of lead fractures.Note that since the electrode array section was not returned, an evaluation and resulting commentary cannot be made on that portion of the lead.
|