MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; OZO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; OZO Back to Search Results
Model Number MMT-751NAL
Device Problems Device Operates Differently Than Expected (2913); Improper Flow or Infusion (2954)
Patient Problems Hyperglycemia (1905); Hypoglycemia (1912); Overdose (1988); Underdose (2542)
Event Date 04/08/2015
Event Type  Injury  
Manufacturer Narrative
A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing. after testing, it was concluded that the device operated within specifications.
 
Event Description
Customer reported via phone call to have high and low blood glucose.Customer's blood glucose ranged from 46 mg/dl to 261mg/dl, which customer treated with glucose tablets.Customer also reported that insulin pump was over delivering and under-delivering.Troubleshooting was performed on insulin pump to determine reason for high blood glucose.After troubleshooting, customer was advised that insulin pump would be replaced due to over delivering.
 
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Manufacturer Narrative
Additional information has been provided by failure analysis that was not included on the initial device evaluation: the insulin pump passed the displacement accuracy test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP
Type of Device
OZO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
christopher tupper
18000 devonshire street
northridge, CA 91325-1219
8185764313
MDR Report Key4784045
MDR Text Key14226798
Report Number3004209178-2015-57765
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
PMA/PMN Number
120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,consumer
Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date 01/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751NAL
Device Catalogue NumberMMT-751NAL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/31/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age53 YR
Patient Weight62
-
-