Model Number MMT-751NAL |
Device Problems
Device Operates Differently Than Expected (2913); Improper Flow or Infusion (2954)
|
Patient Problems
Hyperglycemia (1905); Hypoglycemia (1912); Overdose (1988); Underdose (2542)
|
Event Date 04/08/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing. after testing, it was concluded that the device operated within specifications.
|
|
Event Description
|
Customer reported via phone call to have high and low blood glucose.Customer's blood glucose ranged from 46 mg/dl to 261mg/dl, which customer treated with glucose tablets.Customer also reported that insulin pump was over delivering and under-delivering.Troubleshooting was performed on insulin pump to determine reason for high blood glucose.After troubleshooting, customer was advised that insulin pump would be replaced due to over delivering.
|
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Manufacturer Narrative
|
Additional information has been provided by failure analysis that was not included on the initial device evaluation: the insulin pump passed the displacement accuracy test.
|
|